|
Model 501-B
Smoke Detector Tester
for
Calibrated Sensitivity Inspection
Of
Installed Detectors
Meets Current
NFPA 72-E
UL Listed
Recommendations
For Calibrated
Stable Smoke
Sensitivity
Tests &
Functional Tests
For Ionization
or
Harmless to Detectors
Photoelectric
Detectors
(Duct, wall or
ceiling-mounted)
even older
larger size detectors
30 Seconds Per Test
Adjustable
Range:
Portable, Easy to Operate
0.3 to 4.0
%/ft
Digital Display
National Fire Alarm Code
The National Fire Protection
Association recommends periodic inspection and testing of smoke detectors at
their installed locations as stated by the National Fire Alarm Code (NFPA 72-E,
1996 edition).
The test includes a
calibrated sensitivity test to assure that (1) each detector is within its
listed sensitivity range, and (2) each detector would initiate an alarm when
exposed to smoke of certain concentrations.
The Gemini 501-B Tester
The Gemini 501 Tester combines
the above 2 objectives into a single and easy to perform test. The tester
generates an adjustable amount of aerosol smoke between 0.3 and 4.0 %/ft
obscuration to challenge the detector under test. During the test, the operator
selects a smoke concentration level and directs the smoke through the hand-held
wands and shroud to the detector at the ceiling.
If the concentration selected is below the sensitivity setting of the detector,
the detector would not alarm. However, if the concentration is equal to or
greater than the setting, the detector should alarm within 30 seconds.
The minimum amount of smoke that triggers the alarm is the sensitivity setting
of the detector.
Aerosol Smoke
The tester produces a
consistent and stable smoke from a refined mineral oil. The smoke particle
sizes and size distribution resemble that of the smoke from the UL-268 Standard
Smoke Box. Because of its small particle sizes, low concentration, and short
exposure time, the test smoke is not harmful to the detectors or the operator.
UL Listing and Calibration
Prior to its approval for listing, the
Underwriters laboratories (UL) evaluated to its satisfaction, the correlation of
test results for detector sensitivity between the Gemini 501 Tester and the
UL-268 Standard Smoke Box.
In production, each instrument is individually
calibrated with a Particle Mass Monitor and Standard Reference Detectors. The
instrument output is presented as mass concentration in mg/m3 and
smoke obscuation in %/ft.
Inspection Service
The Gemini Tester was
designed with the inspection personnel in mind. It is a universal tester,
suitable for all types and brands of detectors. Furthermore, it lets you test
the detectors in place, without having to remove them from the ceiling.
Depending on the type of facilities, an experienced operator should be able to
test 150 to 200 detectors per day.
Maintenance
The only regular maintenance
needed is to change the vacuum filter element after 200-300 tests. A pint
bottle of Rudol Liquid, a refined mineral oil, is sufficient for over 800
tests. Standard reference smoke detectors may be used to check the need for
recalibration of the tester.
For detailed specifications
see: Gemini 501B
Brochure.pdf
For prices see:
Gemini Price List
|